Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2018 Apr 6. pii: S1120-1797(18)30057-7. doi: 10.1016/j.ejmp.2018.03.016 |
Quality Controls in Digital Mammography protocol of the EFOMP Mammo Working group. |
Gennaro G1, Avramova-Cholakova S2, Azzalini A3, Luisa Chapel M4, Chevalier M5, Ciraj O6, de Las Heras H7, Gershan V8, Hemdal B9, Keavey E10, Lanconelli N11, Menhart S12, João Fartaria M13, Pascoal A14, Pedersen K15, Rivetti S16, Rossetti V17, Semturs F18, Sharp P19, Torresin A20 |
Abstract This article aims to present the protocol on Quality Controls in Digital Mammography published online in 2015 by the European Federation of Organisations for Medical Physics (EFOMP) which was developed by a Task Force under the Mammo Working Group. The main objective of this protocol was to define a minimum set of easily implemented quality control tests on digital mammography systems that can be used to assure the performance of a system within a set and acceptable range. Detailed step-by-step instructions have been provided, limiting as much as possible any misinterpretations or variations by the person performing. It is intended that these tests be implemented as part of the daily routine of medical physicists and system users throughout Europe in a harmonised way so allowing results to be compared. In this paper the main characteristics of the protocol are illustrated, including examples, together with a brief summary of the contents of each chapter. Finally, instructions for the download of the full protocol and of the related software tools are provided. |
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved. |
KEYWORDS: Digital mammography, EFOMP, Protocol, Quality control |
Publikations ID: 29728229 Quelle: öffnen |